Description
- Very low detection limit – 0.025 µg/L
- High specificity – Insignificant cross-reactivity to C-peptide and proinsulin
- No interference from complex sample matrices
Test Characteristics
Parameter | Value |
Samples | Serum, EDTA plasma, cell culture medium |
Sample volume | 25 µL |
Assay range | 0.025 – 1.0 μg/L |
Detection limit | ≤ 0.025 µg/L |
Incubation | 120 + 30 min |
Specificity
Substance | Cross-reactivity (%) |
Rat insulin | 146 % |
Rat proinsulin | 14 % |
Rat C-peptide | N.D. |
Human insulin | 195 % |
Human proinsulin | 82 % |
Human C-peptide | N.D. |